Radiolucencies surrounding acetabular components with three-dimensional coatings: artifact or real?

2017 
Abstract Background Several 2-dimensional and 3-dimensional surfaces are available for cementless acetabular fixation. Plain radiographs are used to assess osseointegration; however, the radiographs are limited by their inability to capture the bone fixation process occurring over the 3-dimensional cup surface. In this cadaveric study, we compared the bone apposition between 2-dimensional and 3-dimensional cups. Methods Both types of cups were implanted in 6 cadavers and pelvic radiographs obtained. Each cup was resected from the pelvis with adequate bone around it, and subsequently embedded in a polymer. Six sections of each cup were obtained to examine the metal and bone interface. Photographs and contact radiograph images were obtained for each section, and these were graded to arrive at percent metal-bone contact values for the cups. Results On average, Conclusions Good metal-bone contact (75% + contact area) was observed on both cup types. The inconsistencies found in the 3-dimensional cup group may be because of the interaction of radiographs with the unique porous cup surface resulting in artifactual radiolucencies.