Photon-flux determination by the Poisson-fitting technique with quenching corrections

2018 
Abstract Single- and multi-photon spectra of pulsed γ -ray beams produced at 17, 34, and 40 MeV in the laser Compton scattering were measured with an 8” × 12” NaI(Tl) detector. By using the experimental single-photon spectra as the probability function of generating random numbers, response functions of the NaI(Tl) detector to j -fold photons ( j = 2, 3, 4 ...) were constructed. The least-square fits to the experimental multi-photon spectra by the Poisson distribution consisting of the response functions were made. The multi-photon spectra measured at 17 and 34 MeV follow the Poisson distribution. A quenching phenomenon of multi-photon spectra was observed for 40 MeV γ -rays as a result of the saturation at the photomultiplier tube of the NaI(Tl) detector. The original Poisson distributions were restored from the quenched spectra using a saturation curve in the form of y = a ⋅ x η with η =e − b x . We discuss the accuracy of photon-flux determination.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    4
    References
    11
    Citations
    NaN
    KQI
    []