Transiente Elastografie zur Diagnose der Leberzirrhose

2006 
BACKGROUND AND OBJECTIVE: Detection of liver cirrhosis has numerous implications because of the potential sequelae of cirrhosis. Transient elastography (Fibroscan), was evaluated as a novel, non-invasive means of assessing cirrhosis by measuring liver stiffness. METHODS: 147 consecutive patients with different forms of liver disease and histologically determined stages of liver fibrosis were prospectively studied by transient elastography. 48 patients had liver cirrhosis. RESULTS: The number of transient elastographic measurements per patient was 12+/-4 (range 6 - 30). Valid elastography measurements were available for 135 out of 147 patients (92 %). The results of transient elastography correlated positively with the histological score of liver fibrosis (r = 0.8; 95 % CI: 0.72 - 0.85; p or = F3 fibrosis (95 % CI: 0.85 - 0.96) and 0.94 for cirrhosis (95 % CI: 0.90 - 0.98). Using a cut-off value of 13 kPa for detection of liver cirrhosis a sensitivity of 90 %, a specificity of 82 %, a positive predictive value of 71 % and a negative predictive value of 95 % were obtained. CONCLUSIONS: Measuring liver stiffness by transient elastography proved to be an easy method to assess liver cirrhosis. In combination with clinical signs, ultrasound and biochemical markers noninvasive diagnosis of liver cirrhosis will be further improved.