Tissue equivalent proportional counter real time neutron monitor

1978 
The Tissue Equivalent Proportional Counter (TEPC) was developed to monitor low level neutron exposure rates at working stations in a nuclear fuel fabrication facility. It has proven capable of accurately measuring neutron dose rates at levels from 0.1 to 0.2 mrem/hr. It also calculates the Quality Factor which is of importance in locations where the neutron to gamma ratio may vary significantly and irregularily. The system described is computerized to monitor 100 work locations simulationeously and can be expanded to monitor 384 locations. Neutron dose is accumulated on a real time basis and after the proton drop point is established, dose rate can be read out at any time for the dose accumulated over a specified period of time. The current development program which provides reduced system maintenance, lower detection limits, and improved accuracy is discussed along with examples of measurement data and work experience with the current system.