Experimental and numerical investigation of ultrasonic transmission through the skull bone and associated temperature rise

2008 
The feasibility of transcranial high‐intensity focused‐ultrasound (HIFU) therapy within the brain relies on the ability to transmit ultrasound through the skull bone at relatively high ultrasound power. Absorption of ultrasound through the skull bone may cause important temperature rises, and is therefore an important parameter to control. Ultrasonic measurements have shown that the ultrasound beam undergoes a significant attenuation when propagating through the skull, with values on the order of 10 to 20 dB/cm/MHz. To predict temperature rise from such values, it is fundamental to weigh the relative role of absorption to the total ultrasonic attenuation (scattering + absorption + specular reflection). In this work, two types of numerical simulations and experiments are performed to investigate this relative role. Through‐transmission of 1 MHz ultrasound was performed numerically using a 3D Finite‐Difference Time‐Domain (FDTD) algorithm coupled to a 3D bone model obtained from high‐resolution synchrotron ...