Replacement of an aortic Starr-Edwards ball valve prosthesis 28 years after implantation.
2004
A 53-year-old woman who had undergone aortic valve replacement with a Starr-Edwards (S-E) valve (Model 1260) and open mitral commissurotomy 28 years previously was hospitalized with cardiac failure. Echocardiography showed mitral stenosis, mitral regurgitation, and a normally functioning S-E prosthesis. At reoperation, the mitral and aortic valves were replaced with St Jude bileaflet mechanical prostheses. Examination of the explanted S-E prosthesis revealed no structural abnormality other than lipid infiltration of the silastic ball. (Circ J 2004; 68: 507 - 508)