Wearable All-Solid-State Potentiometric Microneedle Patch for Intradermal Potassium Detection

2019 
A new analytical all-solid-state platform for intradermal potentiometric detection of potassium in interstitial fluid is presented here. Solid microneedles are modified with different coatings and polymeric membranes to prepare both the potassium-selective electrode and reference electrode needed for the potentiometric readout. These microneedle-based electrodes are fixed in an epidermal patch suitable for insertion into the skin. The analytical performances observed for the potentiometric cell (Nernstian slope, limit of detection of 10–4.9 potassium activity, linear range of 10–4.2 to 10–1.1, drift of 0.35 ± 0.28 mV h–1), together with a fast response time, adequate selectivity, and excellent reproducibility and repeatability, are appropriate for potassium analysis in interstitial fluid within both clinical and harmful levels. The potentiometric response is maintained after several insertions into animal skin, confirming the resiliency of the microneedle-based sensor. Ex vivo tests based on the intraderm...