Early Flow Disturbances of Tunnelled Haemodialysis Catheters and Topographic Landmarks in Chest X-Ray
2018
Background/Aims: The position of the tip of tunnelled haemodialysis (HD) catheters (THC) might influence flow characteristics during HD. In chest X-ray (CXR), carina-related landmarks may be practicable to verify the THC position, and tip-carina distance (TCD) might be useful to predict early-flow dysfunctions. Methods: In this single-centre, retrospective study, the TCD and the angle between the distal catheter and the body vertical axis (tip-body vertical-angle [TVA]) was measured in 115 THC by post-procedure CXR with 2 investigators. The parameters were proved to be feasible by interrater-reliability and correlated with the incidence of flow-dysfunction within 10 days after insertion. Results: Steep-aligned (TVA p 1.5 cm or left-sighted >4.5 cm below the carina; p Conclusions: Carina-related landmarks in CXR might be helpful to predict early-flow dysfunctions. However, randomized studies will be necessary to confirm this in fluoroscopic-guided placement during the insertion of THC.