Dosimetry errors in endovascular high-dose-rate brachytherapy.
2000
Abstract Monte Carlo data were used to demonstrate the dosimetry of the microSelectron high-dose-rate (HDR) iridium 192 ( 192 Ir) stepping source. These data were used to assess the accuracy of the Nucletron brachytherapy planning system (BPS version 13) for peripheral vessel endovascular brachytherapy. Dose rates from the high-dose-rate (HDR) source are calculated using the Monte Carlo code MCNP4A. Calculations are made at 0.25-cm intervals in the longitudinal direction on sleeves of radii of 1 and 0.25 cm. The Monte Carlo data are summed and weighted to simulate the longitudinal dose distribution at a distance of 1 and 0.25 cm from an 192 Ir source stepping through a straight pathway. A comparison is made between the simulated Monte Carlo dosimetry and the Nucletron brachytherapy planning system’s dosimetry. This study illustrates and quantifies the dosimetric errors at small distances associated with a point source dose calculation algorithm. The effects of step size, dwell time optimization, and active length on the accuracy of BPS v.13 for HDR endovascular brachytherapy are demonstrated.