Automated Attenuation-Based Kilovoltage Selection: Preliminary Observations in Patients After Endovascular Aneurysm Repair of the Abdominal Aorta

2012 
OBJECTIVE. The objective of our study was to assess prospectively the impact of automated attenuation-based kilovoltage selection on image quality and radiation dose in patients undergoing body CT angiography (CTA) after endovascular aneurysm repair (EVAR) of the abdominal aorta. SUBJECTS AND METHODS. Thirty-five patients (five women, 30 men; mean age ± SD, 69 ± 13 years; mean body mass index ± SD, 27.3 ± 4.5 kg/m2) underwent 64-MDCT angiography of the thoracoabdominal aorta using a fixed 120-kVp protocol (scan A: 120 mAs [reference]; rotation time, 0.33 second; pitch, 1.2) and, within a median time interval of 224 days, using a protocol with automated kilovoltage selection (scan B: tube voltage, 80–140 kVp). Subjective image quality (5-point scale: 1 [excellent] to 5 [nondiagnostic]) and objective image quality (aortic attenuation at four locations of the aortoiliac system, noise, contrast-to-noise ratio [CNR]) were assessed independently by two blinded radiologists. The volume CT dose index (CTDIvol) wa...