Sensitive determination of carnosine in urine by high-performance liquid chromatography using 4-(5,6-dimethoxy-2-phthalimidinyl)-2-methoxyphenylsulfonyl chloride as a fluorescent labeling reagent

2010 
Abstract A simple and highly sensitive high-performance liquid chromatography procedure was developed for the determination of carnosine in urine. Carnosine was derivatized with 4-(5,6-dimethoxy-2-phthalimidinyl)-2-methoxyphenylsulfonyl chloride at 70 °C for 15 min in borate buffer (20 mmol l −1 , pH 9.0) to produce fluorescent sulfonamides. After hydrolysis of the reaction mixture with formic acid at 100 °C for 15 min, the fluorescent derivative of carnosine was separated on a reversed-phase column with a linear gradient elution using solvents of (A) acetate buffer (0.1 mmol l −1 , pH 7.0) and (B) acetonitrile at a flow-rate of 1.0 ml/min and was detected at excitation and emission wavelengths of 318 and 400 nm, respectively. The detection limit of carnosine was 4 fmol at a signal-to-noise ratio of 3. The within-day and day-to-day relative standard deviations were 2.7–4.6% and 0.4–5.2%, respectively. The concentration of carnosine in normal human urine was found to be 4.6–125 nmol (mg creatinine) −1 (mean ± SD: 21.6 ± 26.6 nmol (mg creatinine) −1 , n  = 20).
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    13
    Citations
    NaN
    KQI
    []