Blurring the boundaries between frame-based and frameless stereotaxy: feasibility study for brain biopsies performed with the use of a head-mounted robot.

2015 
OBJECT Frame-based stereotactic interventions are considered the gold standard for brain biopsies, but they have limitations with regard to flexibility and patient comfort because of the bulky head ring attached to the patient. Frameless image guidance systems that use scalp fiducial markers offer more flexibility and patient comfort but provide less stability and accuracy during drilling and biopsy needle positioning. Head-mounted robot-guided biopsies could provide the advantages of these 2 techniques without the downsides. The goal of this study was to evaluate the feasibility and safety of a robotic guidance device, affixed to the patient’s skull through a small mounting platform, for use in brain biopsy procedures. METHODS This was a retrospective study of 37 consecutive patients who presented with supratentorial lesions and underwent brain biopsy procedures in which a surgical guidance robot was used to determine clinical outcomes and technical procedural operability. RESULTS The portable head-mount...