Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids

2010 
Background: We evaluated the body fluid (BF) mode on the new Sysmex XE-5000 analyzer. Methods: Red (RBC) and white blood cell (WBC) (differential) counts of BFs (139 patient samples and 87 normal samples) were measured and compared to the Fuchs-Rosenthal chamber and stained cytospin slides. Results: Extended cell counting using the BF mode was noted to have an improved WBC detection limit (CV20%) of 10=10 6 /L. Excellent agreement with the manual method was observed for most BFs wmean bias q 2t o 6=10 6 /L for cerebrospinal fluid (CSF) and –1 to 12=10 6 /L for other fluidsx.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24