Does the fibrin coat around a central venous catheter influence catheter-related sepsis?

1993 
Abstract The purpose of this study was to determine whether the fibrin sheath that develops around an intravenous silicone catheter influences catheter-related sepsis. A rat model with a central intravenous silione catheter was used. Staphylococcus aureus , dose 1 × 10 7 colony forming units (cfu), were injected via the tail vein, either immediately after catheter insertion (group 1, n=23) or after the catheter had been in situ for at least 7 days (group 2, n=22). Blood cultures were done on at 24 hours and 7 days. Animals were killed on day 7 and the catheter was removed for culture (Maki and broth) and scanning electronmicroscopy (SEM). The was no significant difference ( P > .05) between the number of positive blood cultures in groups 1 and 2 at 24 hours (16 v 9) and 7 days (12 v 6). In group 1 there were significantly more positive catheter cultures by both methods (23 v 16 in group 2; P P P = NS). On SEM a fibrinous sheath was identified on all catheters removed on day 7 but not on 5 catheters inserted and removed after 10 minutes. The findings of this study suggest that catheter infection is more likely to occur when bacteremia is introduced immediately following catheter implantation before a fibrinous sheath has developed, compared with delayed bacteremia in the presence of a fibrinous sheath, suggesting a protective role for the sheath.