Portable Pathogen Diagnostics Using Microfluidic Cartridges Made from Continuous Liquid Interface Production Additive Manufacturing.

2021 
Biomedical diagnostics based on microfluidic devices have the potential to significantly benefit human health; however, the manufacturing of microfluidic devices is a key limitation to their widespread adoption. Outbreaks of infectious disease continue to demonstrate the need for simple, sensitive, and translatable tests for point-of-care use. Additive manufacturing (AM) is an attractive alternative to conventional approaches for microfluidic device manufacturing based on injection molding; however, there is a need for development and validation of new AM process capabilities and materials that are compatible with microfluidic diagnostics. In this paper, we demonstrate the development and characterization of AM cartridges using continuous liquid interface production (CLIP) and investigate process characteristics and capabilities of the AM microfluidic device manufacturing. We find that CLIP accurately produces microfluidic channels as small as 400 μm and that it is possible to routinely produce fluid channels as small as 100 μm with high repeatability. We also developed a loop-mediated isothermal amplification (LAMP) assay for detection of E. coli from whole blood directly on the CLIP-based AM microfluidic cartridges, with a 50 cfu/μL limit of detection, validating the use of CLIP processes and materials for pathogen detection. The portable diagnostic platform presented in this paper could be used to investigate and validate other AM processes for microfluidic diagnostics and could be an important component of scaling up the diagnostics for current and future infectious diseases and pandemics.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    50
    References
    1
    Citations
    NaN
    KQI
    []