Evaluation of respiratory motion-corrected cone-beam CT at end expiration in abdominal radiotherapy sites: a prospective study

2018 
AbstractBackground: Cone beam computed tomography (CBCT) for radiotherapy image guidance suffers from respiratory motion artifacts. This limits soft tissue visualization and localization accuracy, particularly in abdominal sites. We report on a prospective study of respiratory motion-corrected (RMC)-CBCT to evaluate its efficacy in localizing abdominal organs and improving soft tissue visibility at end expiration.Material and methods: In an IRB approved study, 11 patients with gastroesophageal junction (GEJ) cancer and five with pancreatic cancer underwent a respiration-correlated CT (4DCT), a respiration-gated CBCT (G-CBCT) near end expiration and a one-minute free-breathing CBCT scan on a single treatment day. Respiration was recorded with an external monitor. An RMC-CBCT and an uncorrected CBCT (NC-CBCT) were computed from the free-breathing scan, based on a respiratory model of deformations derived from the 4DCT. Localization discrepancy was computed as the 3D displacement of the GEJ region (GEJ patie...