The impact of different voxels and exposure parameters of CBCT for the assessment of external root resorptions: A phantom study

2019 
: This study aimed to compare the effectiveness of the different voxel sizes and imaging parameters of cone beam computed tomography (CBCT) on the determination of external root resorption (ERR). ERRs were created in coronal, middle and apical regions. Depths (0.62-0.73 mm) and diameters (0.68-0.81 mm) of ERRs were measured using a digital calliper. Images were obtained in 0.125, 0.2, 0.25, 0.3 and 0.4 mm3 voxel sizes and at different currents. Data were analysed using Friedman and Wilcoxon signed-rank tests at a level of P   0.05). The change in current and exposure time did not make a difference in determining the dimensions of the ERRs (P > 0.05). Some voxel dimensions were more reliable in determining the depth and diameter of ERR in the coronal and middle region.