Validation of arteriograph - A new oscillometric device to measure arterial stiffness in patients on maintenance hemodialysis

2009 
Background: Measuring arterial stiffness (augmentation index (AI), aortic pulse wave velocity (PWV)) in hemodialysis (HD) patients has prognostic significance. To assess its validity, the new oscillometric Arteriograph device (AIA, PWVA) was compared to the validated PulsePen tonometer (AIP, PWVP). Methods: AI and PWV were measured in 98 patients with both devices before HD. Validity was evaluated by Pearson’s correlation, Bland-Altman analysis, and by assessing the prognostic value of AI and PWV to predict cardiovascular (CV) mortality over 29 months. Results: Correlation between AIP and AIA was significant (R = 0.527, p P and PWVA readings (R = 0.173, p = 0.097). The average difference of PWV values by the two devices was –1.2 m/s, and 20.6% of the paired PWV differences fall outside the ±1 SD boundary. In survival analyses, only PWVP but not PWVA was significantly related to CV mortality. Conclusion: Lack of correlation between PWVP and PWVA and lack of prognostic significance of PWVA suggest limited validity of Arteriograph to determine PWV in patients on HD.