Cassette incubation followed by bioanalysis using high-resolution MS for in vitro ADME screening assays

2012 
Background: High-resolution MS (HRMS) has recently received a considerable interest in quantitative bioanalysis using full-scan acquisition mode. The benefits include complete elimination of compound-specific MS method development, and simultaneous collection of mass spectral data on both targeted and non-targeted components. One additional advantage that has not been widely discussed is its suitability for simultaneous quantitation of, theoretically, an unlimited number of compounds, which is not possible with selected reaction monitoring (SRM) on a triple quadrupole mass spectrometer. Materials & Methods: We took advantage of this unique bioanalytical capability of HRMS and developed a novel in vitro ADME workflow of cassette incubation of as many as 32 compounds, followed by quantitative bioanalysis using full-scan acquisition on an Orbitrap HRMS. The workflow was evaluated for a serum protein-binding assay and a parallel artificial membrane permeability (PAMPA) assay. Results: The bioanalytical assay ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    22
    References
    23
    Citations
    NaN
    KQI
    []