Internal Quality Control Practices in Coagulation Laboratories: recommendations based on a patterns‐of‐practice survey

2015 
SummaryIntroduction Internal quality control (IQC) procedures are crucial for ensuring accurate patient test results. The IQMH Centre for Proficiency Testing conducted a web-based survey to gather information on the current IQC practices in coagulation testing. Methods A questionnaire was distributed to 174 Ontario laboratories licensed to perform prothrombin time (PT) and activated partial thromboplastin time (APTT). Results All laboratories reported using two levels of commercial QC (CQC); 12% incorporate pooled patient plasma into their IQC program; >68% run CQC at the beginning of each shift; 56% following maintenance, with reagent changes, during a shift, or with every repeat sample; 6% only run CQC at the beginning of the day and 25% when the instruments have been idle for a defined period of time. IQC run frequency was determined by manufacturer recommendations (71%) but also influenced by the stability of test (27%), clinical impact of an incorrect test result (25%), and sample's batch number (10%). IQC was monitored using preset limits based on standard deviation (66%), precision goals (46%), or allowable performance limits (36%). 95% use multirules. Failure actions include repeating the IQC (90%) and reporting patient results; if repeat passes, 42% perform repeat analysis of all patient samples from last acceptable IQC. Conclusion Variability exists in coagulation IQC practices among Ontario clinical laboratories. The recommendations presented here would be useful in encouraging standardized IQC practices.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    1
    Citations
    NaN
    KQI
    []