The Performance of LiF:Mg-Ti for Proton Dosimetry within the Framework of the MoVe IT Project

2021 
Proton therapy represents a technologically advanced method for delivery of radiation treatments to tumors. The determination of the biological effectiveness is one of the objectives of the MoVe IT (Modeling and Verification for Ion Beam Treatment Planning) project of the National Institute for Nuclear Physics (INFN) CSN5. The aim of the present work, which is part of the project, was to evaluate the performance of the thermoluminescent dosimeters (TLDs-100) for dose verification in the proton beam line. Four irradiation experiments were performed in the experimental room at the Trento Proton Therapy Center, where a 150 MeV monoenergetic proton beam is available. A total of 80 TLDs were used. The TLDs were arranged in one or two rows and accommodated in a specially designed water-equivalent phantom. In the experimental setup, the beam enters orthogonally to the dosimeters and is distributed along the proton beam profile, while the irradiation delivers doses of 0.8 Gy or 1.5 Gy in the Bragg peak. For each irradiation stage, the depth–dose curve was determined by the TLD readings. The results showed the good performance of the TLDs-100, proving their reliability for dose recordings in future radiobiological experiments planned within the MoVe IT context.