Evaluation of a New Handheld Instrument for the Detection of Counterfeit Artesunate by Visual Fluorescence Comparison

2014 
There is an urgent need for accurate and inexpensive handheld instruments for the evaluation of medicine quality in the field. A blinded evaluation of the diagnostic accuracy of the Counterfeit Detection Device 3 (CD-3), developed by the US Food and Drug Administration Forensic Chemistry Center, was conducted in the Lao People's Democratic Republic. Two hundred three samples of the oral antimalarial artesunate were compared with authentic products using the CD-3 by a trainer and two trainees. The specificity (95% confidence interval [95% CI]), sensitivity (95% CI), positive predictive value (95% CI), and negative predictive value (95% CI) of the CD-3 for detecting counterfeit (falsified) artesunate were 100% (93.8–100%), 98.4% (93.8–99.7%), 100% (96.2–100%), and 97.4% (90.2–99.6%), respectively. Interobserver agreement for 203 samples of artesunate was 100%. The CD-3 holds promise as a relatively inexpensive and easy to use instrument for field evaluation of medicines, potentially empowering drug inspectors, customs agents, and pharmacists.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    39
    Citations
    NaN