Artefacts in multimodal imaging of titanium, zirconium and binary titanium-zirconium alloy dental implants: an in vitro study.

2017 
Objectives:To analyze and evaluate imaging artefacts induced by zirconium, titanium and titaniumzirconium alloy dental implants.Methods:Zirconium, titanium and titaniumzirconium alloy implants were embedded in gelatin and MRI, CT and CBCT were performed. Standard protocols were used for each modality. For MRI, line–distance profiles were plotted to quantify the accuracy of size determination. For CT and CBCT, six shells surrounding the implant were defined every 0.5 cm from the implant surface and histogram parameters were determined for each shell.Results:While titanium and titaniumzirconium alloy induced extensive signal voids in MRI owing to strong susceptibility, zirconium implants were clearly definable with only minor distortion artefacts. For titanium and titaniumzirconium alloy, the MR signal was attenuated up to 14.1 mm from the implant. In CT, titanium and titaniumzirconium alloy resulted in less streak artefacts in comparison with zirconium. In CBCT, titaniumzirconium alloy induced more s...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    38
    References
    28
    Citations
    NaN
    KQI
    []