Comparison of Elastic and Rigid Registration during Magnetic Resonance Imaging/Ultrasound Fusion-Guided Prostate Biopsy: A Multi-Operator Phantom Study

2018 
Purpose: The relative value of rigid or elastic registration during magnetic resonance imaging/ultrasound fusion guided prostate biopsy has been poorly studied. We compared registration errors (the distance between a region of interest and fiducial markers) between rigid and elastic registration during fusion guided prostate biopsy using a prostate phantom model.Materials and Methods: Four gold fiducial markers visible on magnetic resonance imaging and ultrasound were placed throughout 1 phantom prostate model. The phantom underwent magnetic resonance imaging and the fiducial markers were labeled as regions of interest. An experienced user and a novice user of fusion guided prostate biopsy targeted regions of interest and then the corresponding fiducial markers on ultrasound after rigid and then elastic registration. Registration errors were compared.Results: A total of 224 registration error measurements were recorded. Overall elastic registration did not provide significantly improved registration error...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    25
    References
    11
    Citations
    NaN
    KQI
    []