Automated Gas-Phase Purification for Accurate, Multiplexed Quantification on a Stand-Alone Ion-Trap Mass Spectrometer

2013 
Isobaric tagging enables the acquisition of highly multiplexed proteome quantification, but it is hindered by the pervasive problem of precursor interference. The elimination of coisolated contaminants prior to reporter tag generation can be achieved through the use of gas-phase purification via proton transfer ion/ion reactions (QuantMode); however, the original QuantMode technique was implemented on the high-resolution linear ion-trapOrbitrap hybrid mass spectrometer enabled with electron transfer dissociation (ETD). Here we extend this technology to stand-alone linear ion-trap systems (trapQuantMode, trapQM). Facilitated by the use of inlet beam-type activation (i.e., trapHCD) for production and observation of the low mass-to-charge reporter region, this scan sequence comprises three separate events to maximize peptide identifications, minimize duty cycle requirements, and increase quantitative accuracy, precision, and dynamic range. Significant improvements in quantitative accuracy were attained over...