Lossy (15:1) JPEG Compression of Digital Coronary Angiograms Does Not Limit Detection of Subtle Morphological Features

1996 
Background Development of the “all-digital” cardiac catheterization laboratory has been slowed by substantial computer archival and transfer requirements. Lossy data compression reduces this burden but creates irreversible changes in images, potentially impairing detection of clinically important angiographic features. Methods and Results Fifty image sequences from 31 interventional procedures were viewed both in the original (uncompressed) state and after 15:1 lossy Joint Photographic Expert’s Group (JPEG) compression. Experienced angiographers identified dissections, suspected thrombi, and coronary stents, and their results were compared with those from a consensus panel that served as a “gold standard.” The panel and the individual observers reviewed the same image sequences 4 months after the first session to determine intraobserver variability. Intraobserver agreement for original images was not significantly different from that for compressed images (89.8% versus 89.5% for 600 pairs of observations ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    20
    References
    31
    Citations
    NaN
    KQI
    []