Analytical validation of a highly sensitive point-of-care system for cardiac troponin I determination

2019 
Background Highly sensitive cardiac troponin assays (hs-cTn) are not available as point-of-care (POC) measurements. As rapid testing cannot be achieved at the expense of clinical performance, there is an urgent need to develop and rigorously validate POC hs-cTn. Konica Minolta (KM) has recently developed a surface plasmon-field enhanced fluorescence spectroscopy-based POC hs-cTn I system. Methods We validated the analytical characteristics of the KM POC system according to the international guidelines. Results Limit of blank (LoB) and limit of detection (LoD) were 0.35 and 0.62 ng/L, respectively, hs-cTn I concentrations corresponding to a total CV of 20%, 10% and 5% were 1.5, 3.9 and 11.0 ng/L, respectively. Method comparison studies showed that KM calibration was successfully traced to higher-order references. Limit of quantitation (LoQ), i.e. the hs-cTn I concentration having a total error of measurement of LoD in 65.7% of all reference individuals, in 76.7% of males and in 54.7% of females, respectively. Conclusions The KM system joins the characteristics of POC systems to the analytical performance of hs-cTn.