Two Serum Free Light Chain Detection Systems in the Diagnosis of Multiple Myeloma

2021 
Objective To investigate the comparability of the Freelite, Binding Site, Beckman and N Latex FLC, Siemens in the detection of serum free light chain (sFLC) . Methods Fifty newly diagnosed multiple myeloma (MM) patients in Tianjin Institute of Blood Research from November 2019 to February 2020 were enrolled. The two systems (Freelite, Binding Site, Beckman and N Latex FLC, Siemens) were used to detect the sFLC of the samples. Outlier detection was performed by ESD method, methodological comparison and deviation assessment were performed by Passing-Bablok regression and Bland-Altman regression. Results Both the systems could quantitatively analyze free kappa light chain serum samples and free lambda light chain samples. Freelite, Binding Site, Beckman and N Latex FLC, Siemens free light chain test showed FLC-κ:36.5 (6.5, 194), 40.5 (6.94, 288), FLC-λ: 30.1 (4.3, 170.5), 35.1 (2.28, 526), rFLC (FLC-κ/ FLC-λ) : 0.82 (0.05, 43.25), 1.03 (0.03, 32.04), dFLC (|FLC-κ- FLC-λ|) : -5.8 (-161.97, 183.7), 1.1 (-505.1, 279.01), which existed no outliers. There were systematic differences, and the deviation level was not within the clinically acceptable range. Conclusion Both the systems can meet the needs of clinical diagnosis and treatment, but there is a significant deviation between the two systems, the results are not comparable, and should be analyzed separately. In particular, the same system should be selected for monitoring the prognosis of MM.