Evaluation of different parameters affecting the liquid scintillation spectrometry measurement of gross alpha and beta index in water samples

2011 
Abstract Liquid scintillation spectrometry is a fast competitive technique for the simultaneous evaluation of gross alpha and beta indexes. However, the implementation of this technique should not be considered as straightforward, and the pre-concentration methods to decrease the detection limit together with quenching and alpha, and beta crossover corrections should be carefully chosen according to the needs of the laboratory. Both aspects are being approached in this work as to find an easy and robust method for alpha/beta measurement in water samples, taking into account the quenching and alpha/beta crossover interferences effects. Results showed that most of the pre-concentration methods increased the quenching in the measurement, although HNO 3 0.05 M points to be the best solution for pre-concentration and re-dissolution of the sample as converges into low quenching and maximum recovery. Subsequently, in the measurement of water samples with different conductivities, the analysis of the raw counts to obtain gross alpha and beta indexes was carried out using different approaches to implement quenching and interference corrections. If quenching and salt content in the sample are relatively low, interference and quenching-efficiency corrections do not improve the accuracy of the results within the usual precision assumed for a result of gross alpha and beta index (25%). Special attention must be paid when corrections are applied to high quenched or saline samples and when alpha and beta activities values are different in several orders of magnitude.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    13
    Citations
    NaN
    KQI
    []