PA/Lateral Chest X‐Ray is Equivalent to Cine‐Fluoroscopy for the Detection of Conductor Externalization in Defibrillation Leads
2015
Background
Riata™ and Riata ST defibrillation leads (St. Jude Medical, Sylmar, CA, USA) are susceptible to insulation defects with conductor externalization. Cine-fluoroscopy is considered to be the gold standard for the documentation of insulation defects, but similar detection rates have been reported for posterior-anterior (PA)/lateral chest x-ray (CXR) with zooming.
Objective
Prospective single-center study to assess the diagnostic equivalence of a PA/lateral CXR with zooming for the detection of Riata insulation defects in a direct comparison to cine-fluoroscopy.
Methods
Seventy-eight consecutive patients underwent 3-view cine-fluoroscopy and a PA/lateral CXR. All CXRs and cine-fluoroscopy images were reviewed by blinded electrophysiologists and staff radiologists.
Results
Forty-four of 78 patients had an abnormal cine-fluoroscopy (56%). The diagnostic correlation between PA/lateral CXR and cine-fluoroscopy was excellent (κ = 0.90; 95% confidence interval 0.80–1.00). PA/lateral CXR was equivalent to cine-fluoroscopy for the detection of conductor externalization showing a sensitivity of 97.7% and a specificity of 91.2%. The mean radiation effective dose of CXR was significantly lower compared to cine-fluoroscopy (0.09 millisievert [mSV] vs 0.85 ± 0.47 mSv; P < 0.001). Also, CXR was significantly less expensive.
Conclusion
PA/lateral CXR with zooming is equivalent to cine-fluoroscopy for the detection of Riata insulation defects and should be considered as the preferred screening method.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
11
References
1
Citations
NaN
KQI