Comparison of cardiac TnI outliers using a contemporary and a high-sensitivity assay on the Abbott Architect platform

2014 
BackgroundAssays for cardiac troponin (cTn) have undergone improvements in sensitivity and precision in recent years. Increased rates of outliers, however, have been reported on various cTn platforms, typically giving irreproducible, falsely higher results. We aimed to evaluate the outlier rate occurring in patients with elevated cTnI using a contemporary and high-sensitivity assay.MethodsAll patients with elevated cTnI (up to 300 ng/L) performed over a 21-month period were assayed in duplicate. A contemporary assay (Abbott STAT Troponin-I) was used for the first part of the study and subsequently a high-sensitivity assay (Abbott STAT High-Sensitive Troponin-I) was used. Outliers exceeded a calculated critical difference (CD) (CD = z × √2 × SDAnalytical) where z = 3.5 (for probability of 0.0005) and critical outliers also were on a different side of the decision level.ResultsThe respective outlier and critical outlier rates were 0.22% and 0.10% for the contemporary assay (n = 4009) and 0.18% and 0.13% for...