Low-energy in-line electron holographic imaging of vitreous ice-embedded small biomolecules using a modified scanning electron microscope

2020 
Abstract Cryo-electron microscopy (cryo-EM) has become the method of choice in the field of structural biology, owing to its unique ability to deduce structures of vitreous ice-embedded, hydrated biomolecules over a wide range of structural resolutions. As cryo-transmission electron microscopes (cryo-TEM) become increasingly specialised for high, near-atomic resolution studies, operational complexity and associated costs serve as significant barriers to widespread usability and adoptability. To facilitate the expansion and accessibility of the cryo-EM method, an efficient, user-friendly means of imaging vitreous ice-embedded biomolecules has been called for. In this study, we present a solution to this issue by integrating cryo-EM capabilities into a commercial scanning electron microscope (SEM). Utilising the principle of low-energy in-line electron holography, our newly developed hybrid microscope permits low-to-moderate resolution imaging of vitreous ice-embedded biomolecules without the need for any form of sample staining or chemical fixation. Operating at 20 kV, the microscope takes advantage of the ease-of-use of SEM-based imaging and phase contrast imaging of low-energy electron holography. This study represents the first reported successful application of low-energy in-line electron holographic imaging to vitreous ice-embedded small biomolecules, the effectiveness of which is demonstrated here with three morphologically distinct specimens.