Implant preloading in extension reduces spring length change in dynamic intraligamentary stabilization: a biomechanical study on passive kinematics of the knee
2018
Purpose
Dynamic intraligamentary stabilization (DIS) is a primary repair technique for acute anterior cruciate ligament (ACL) tears. For internal bracing of the sutured ACL, a metal spring with 8 mm maximum length change is preloaded with 60–80 N and fixed to a high-strength polyethylene braid. The bulky tibial hardware results in bone loss and may cause local discomfort with the necessity of hardware removal. The technique has been previously investigated biomechanically; however, the amount of spring shortening during movement of the knee joint is unknown. Spring shortening is a crucial measure, because it defines the necessary dimensions of the spring and, therefore, the overall size of the implant.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
25
References
5
Citations
NaN
KQI