Evidence for age-related performance degradation of 241Am foil sources commonly used in UK schools

2014 
The characteristics of alpha radiation have for decades been demonstrated in UK schools using small sealed 241Am sources. There is a small but steady number of schools who report a considerable reduction in the alpha count rate detected by an end-window GM detector compared with when the source was new. This cannot be explained by incorrect apparatus or set-up, foil surface contamination, or degradation of the GM detector. The University of Liverpool and CLEAPSS collaborated to research the cause of this performance degradation. The aim was to find what was causing the performance degradation and the ramifications for both the useful and safe service life of the sources. The research shows that these foil sources have greater energy straggling with a corresponding reduction in spectral peak energy. A likely cause for this increase in straggling is a significant diffusion of the metals over time. There was no evidence to suggest the foils have become unsafe, but precautionary checks should be made on old sources.