Utility of viscoelastography with TEG6s to direct management of haemostasis during obstetric haemorrhage: a prospective observational study

2021 
Abstract Background The TEG 6s is an automated cartridge-based device with limited description of use in obstetric haemorrhage. The aim of this analysis was to describe the utility of TEG 6s in identifying abnormal laboratory results of coagulation and platelet count, and inform an interventional treatment algorithm for postpartum haemorrhage. Methods A prospective observational cohort study of 521 women with moderate to severe obstetric haemorrhage (>1000 mL blood loss), including 372 women with at least one TEG 6s test. A non-pregnant control group was used for reference. TEG 6s test parameters Citrated Functional Fibrinogen (CFF), Citrated Kaolin TEG (CK) and Citrated Rapid TEG (CRT) were compared with paired laboratory tests of fibrinogen, PT/aPTT and platelet count, obtained during haemorrhage. Results Among 456 TEG 6s tests, 389 were matched with laboratory coagulation results. The receiver operator characteristic area-under-the-curve (95% CI) for CFF amplitude by 10 min to detect Clauss fibrinogen ≤2 g/L was 0.95 (0.91 to 0.99) (P Conclusion The TEG 6s CFF can be used to identify low fibrinogen during obstetric haemorrhage. A value to identify transfusion thresholds for PT/aPTT and platelets was not established, and laboratory results should continue to be used.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    23
    References
    1
    Citations
    NaN