ASSESSMENT OF CUP ORIENTATION IN HIP RESURFACING: A COMPARISON OF TRAUMACAD AND COMPUTED TOMOGRAPHY

2018 
The orientation of the acetabular component in metal-on-metal hip resurfacing arthroplasty affects wear rate and hence failure. Correct assessment of acetabular orientation is key in assessing the painful hip resurfacing. This study aimed to establish if interpretation of pelvic radiographs with TraumaCad software can provide a reliable alternative to computed tomography (CT) in measuring the acetabular inclination and version.TraumaCad was used to measure the acetabular orientation on AP pelvis radiographs of 14 painful hip resurfacings. Four orthopaedic surgeons performed each measurement twice. These were compared with measurements taken from CT reformats performed by an experienced musculoskeletal radiologist. The correlation between TraumaCad and CT was calculated, as was the intra- and inter-observer reliability of TraumaCad.There is strong correlation between the two techniques for the measurement of inclination and version (p<0.001). Intra- and inter-observer reliability of TraumaCad measurements ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []