Performance evaluation of the ADVIA Centaur anti-HBe and HBeAg assays.

2008 
Abstract Background Detection of HBeAg and anti-HBe is valuable for the evaluation and therapeutic management of hepatitis B infection. Objectives To determine the clinical performance of the newly CE-approved a aAvailable outside the US. FDA clearance pending. HBeAg and anti-HBe assays on the fully automated, random access ADVIA Centaur ® immunoassay system. Study design Patient samples collected at two sites were used to compare the ADVIA Centaur assays to Abbott AxSYM™ assays. Consensus of discordant results was reached using Roche Elecsys ® assays. Additionally, two well-characterized seroconversion panels were evaluated. Results The ADVIA Centaur HBeAg assay sensitivity was 100% and specificity was 99.5%. The ADVIA Centaur anti-HBe assay sensitivity was 100% and the resolved specificity was 98.2%. Fewer samples required retesting with the ADVIA Centaur assays than with the AxSYM. In two well-characterized seroconversion panels, the ADVIA Centaur anti-HBe assay detected anti-HBe 20–25 days earlier than the AxSYM assay; the ADVIA Centaur and AxSYM HBeAg assays detected HBe reactivity on the same day. Conclusions The ADVIA Centaur HBeAg and anti-HBe assays demonstrated good sensitivity and specificity, and thus are suitable for clinical use. Their novel algorithms require reduced retesting, suggesting these assays may be more cost effective.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    9
    References
    5
    Citations
    NaN
    KQI
    []