Spirometry with bronchodilator test: effect that the use of large-volume spacers with antistatic treatment has on test response* Prova broncodilatadora na espirometria: efeito do uso de espaçador de grande volume com tratamento antiestático na resposta ao broncodilatador

2011 
Objective: To evaluate whether the use of inhaled albuterol via a metered-dose inhaler with a large-volume spacer with antistatic treatment modifies the bronchodilator test results when compared with the usual technique (no spacer). Methods: A prospective study involving 24 patients, 18-45 years of age, clinically suspected of having asthma, and under treatment at the Outpatient Pulmonary Clinic of the Federal University of Minas Gerais Hospital das Clinicas, located in the city of Belo Horizonte, Brazil. All of the patients underwent two bronchodilator tests: one with and one without the use of a large-volume spacer. Results: There was no significant difference in the variation of FEV 1 prior to and after bronchodilator use between the two techniques (mean ΔFEV 1 = 0.01 L; 95% CI: −0.05 to 0.06; p = 0.824). No statistically significant difference was found between the two techniques regarding the qualitative results on the bronchodilator test (p = 1.00). There was concordance between the techniques in terms of the bronchodilator test results (kappa coefficient = 0.909; p < 0.005). Conclusions: According to the results of this study, the use of large-volume spacers does not significantly modify bronchodilator test results.