Purpose: A cochlea modeling approach is presented allowing for a user defined degree of geometry simplification which automatically adjusts to the patient specific anatomy. Model generation can be performed in a straightforward manner due to error estimation prior to the actual generation, thus minimizing modeling time. Therefore, the presented technique is well suited for a wide range of applications including finite element analyses where geometrical simplifications are often inevitable. Methods: The method is presented for n=5 cochleae which were segmented using a custom software for increased accuracy. The linear basilar membrane cross sections are expanded to areas while the scalae contours are reconstructed by a predefined number of arc segments. Prior to model generation, geometrical errors are evaluated locally for each cross section as well as globally for the resulting models and their basal turn profiles. The final combination of all reconditioned features to a 3D volume is performed in Autodes...