Standardization of a Bioassay for Thyrotropin Receptor Stimulating Autoantibodies

2015 
Background: Cell-based bioassays for functional thyroid stimulating autoantibodies (TSAb) are sensitive diagnostic tools. However, there is no bioassay available that is standardized with international reference material. We aimed to promote the standardization of the test results among laboratories that perform TSAb bioassays and calibrate TSAb levels against the second international standard (IS) 08/204 from the National Institute for Biological Standards and Control (NIBSC). Methods: Serum TSAb activity was measured with a FDA-cleared bioassay that utilizes CHO cells expressing a chimeric thyrotropin receptor (TSHR) and a c-AMP response-element-dependent luciferase. The IS was applied for calibration. TSAb results were reported as percentage of specimen-to-reference ratio (SRR%) and converted into mIU/L. Results: The IS dose–response curve was obtained using concentrations from 0.3125 to 200 mIU/L. Mean TSAb SRR%±standard deviation (SD) values for the IS concentrations 0.3125, 0.625, 1.25, 2.5, 5, 10, ...