Comparison of Nikkiso and Bio-Medicus Pumps in Thoracoabdominal Aortic Surgery

1999 
Left heart bypass reduces the risk of ischemic complications during the repair of extensive thoracoabdominal aortic aneurysms. This prospective study compared the performance of the recently developed Nikkiso centrifugal pump with the well-established Bio-Medicus pump during left heart bypass for thoracoabdominal aortic aneurysm surgery. Thirty-five consecutive patients undergoing graft repair of extensive thoracoabdominal aortic aneurysms were prospectively assigned to have left heart bypass using either the Bio-Medicus (in the first 19 patients) or Nikkiso pump (in the next 16 patients). There were no significant differences in pump flow rates or patient hemodynamics between the two groups and there was no evidence of pump malfunction. All patients survived and none developed postoperative coagulopathy, myocardial infarction, or left heart failure. Paraparesis developed in 2 patients in the Nikkiso group (12.5%); there were no neurologic complications in the Bio-Medicus group (p = 0.202). One patient in...