Characterization and compensation of f0 inhomogeneity artifact in spiral hyperpolarized 13 C imaging of the human heart.

2021 
PURPOSE This study aimed to investigate the role of regional f0 inhomogeneity in spiral hyperpolarized 13 C image quality and to develop measures to alleviate these effects. METHODS Field map correction of hyperpolarized 13 C cardiac imaging using spiral readouts was evaluated in healthy subjects. Spiral readouts with differing duration (26 and 45 ms) but similar resolution were compared with respect to off-resonance performance and image quality. An f0 map-based image correction based on the multifrequency interpolation (MFI) method was implemented and compared to correction using a global frequency shift alone. Estimation of an unknown frequency shift was performed by maximizing a sharpness objective based on the Sobel variance. The apparent full width half at maximum (FWHM) of the myocardial wall on [13 C]bicarbonate was used to estimate blur. RESULTS Mean myocardial wall FWHM measurements were unchanged with the short readout pre-correction (14.1 ± 2.9 mm) and post-MFI correction (14.1 ± 3.4 mm), but significantly decreased in the long waveform (20.6 ± 6.6 mm uncorrected, 17.7 ± 7.0 corrected, P = .007). Bicarbonate signal-to-noise ratio (SNR) of the images acquired with the long waveform were increased by 1.4 ± 0.3 compared to those acquired with the short waveform (predicted 1.32). Improvement of image quality was observed for all metabolites with f0 correction. CONCLUSIONS f0 -map correction reduced blur and recovered signal from dropouts, particularly along the posterior myocardial wall. The low image SNR of [13 C]bicarbonate can be compensated with longer duration readouts but at the expense of increased f0 artifacts, which can be partially corrected for with the proposed methods.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    25
    References
    2
    Citations
    NaN
    KQI
    []