Compressibility Effects in Packed and Open Tubular Gas and Supercritical Fluid Chromatography
1992
Summary The influence of the pressure drop on the efficiency and speed of analysis in packed and open tubular supercritical fluid chromatography (SFC) is described: methods previously developed to describe the effects of mobile phase compressibility on the performance of open tubular columns in SFC have been extended to packed columns. The Horvath and Lin equation has been used to elucidate the influence of variations in velocity, diffusivity, and capacity factor along the column on the overall efficiency of packed column SFC. In packed columns, in contrast with the situation in open tubular columns, because the increase in velocity is no longer compensated by an increase in diffusion coefficients, the increase in both linear velocity and capacity factor which result from a significant pressure drop cause the plate height to increase along the column. The effect of fluid decompression along the length of the column on the speed of analysis in SFC has been studied and numerical expressions derived which enable calculation of compressibility correction factors for the plate height. Both the fl and f2 correction factors remain very close to unity for acceptable pressure drops, which means that the pressure drop has virtually no effect on the number of plates generated per unit time for an unretained component. For retained species, the decompression of the mobile phase across the column causes the capacity factor to increase and hence leads to increased analysis times.