This study compares the histologic effects of scalpel, CO2 laser, electrosurgery, and constant-voltage electrosurgery incisions on the mucosal tissue of swine. Tissue studies comparing the CO2 laser with the scalpel and electrosurgery have been done. However, a gross and histologic comparison of the effects of all three techniques on oral mucosal tissue has not been reported. A swine model of both tongue and buccal mucosa was used to compare the scalpel, CO2 laser, electrosurgery unit, and constant-voltage electrosurgery unit in an effort to assess their value in oral surgery. Tissue samples of tongue and buccal mucosal incisions and excisions were histologically examined at 0, 3, 7, 14, 28, and 42 days after surgery to evaluate tissue damage and wound healing properties induced by the four instruments. The instruments were also evaluated for performance and ease of use. On subjective evaluation of ease of use, constant-voltage electrosurgery scored highest (p < 0.05) on a scale of 0 to 4, followed by the...