Measurement of the weighted peak level for occupational exposure to gradient magnetic fields for 1.5 and 3 Tesla MRI body scanners

2015 
The purpose of this work is to give a contribution to the construction of a comprehensive knowledge of the exposure levels to gradient magnetic fields (GMF) in terms of the weighed peak (WP), especially for 3 Tesla scanners for which there are still few works available in the literature. A new generation probe for the measurement of electromagnetic fields in the range of 1 Hz-400 kHz was used to assess the occupational exposure levels to the GMF for 1.5 and 3.0 Tesla MRI body scanners, using the method of the WP according to the International Commission on Non-Ionizing Radiation Protection (ICNIRP) approach. The probe was placed at a height of 1.1 m, close to the MRI scanners, where operators could stay during some medical procedures with particular issues. The measurements were performed for a set of typical acquisition sequences for body (liver) and head exams. The measured values of WP were in compliance with ICNIRP 2010 reference levels for occupational exposures.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    11
    References
    5
    Citations
    NaN