Radiofrequency perforation of the pulmonary valve: an efficient low cost solution

2017 
AbstractObjective The aim of the study was to assess the feasibility of using commonly available catheterization laboratory equipment for radiofrequency perforation of the pulmonary valve in patients with pulmonary atresia and intact ventricular septum.Methods The system (off-label use for all items) is made up of a co-axial telescopic arrangement consisting of a 0.014" PT2 ™ coronary guidewire, for insulation inside a 2.7-F microcatheter which has an inner lumen of 0.021". The microcatheter was passed via a standard 4-F right coronary catheter to just below the atretic pulmonary valve. Radiofrequency (RF) energy was delivered using a standard electrosurgical system. In vitro testing had been performed and indicated that 5-10 W for 2-5 s would be sufficient for valve perforation.Results Radiofrequency perforation was successfully performed in all (n = 5, 100%) patients at a median age of 3 days (range: 1-36) and weight 2.7 kg (range 2.3-3.0). In one patient the pericardium was entered during the initial a...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    14
    References
    1
    Citations
    NaN
    KQI
    []