Overcoming the Hurdle of Fluorescent Compounds in Kinase Screening: A Case Study

2006 
In the field of screening in general and especially in the kinase area, taking into consideration throughput and cost, fluorescence- and luminescence-based assays have been developed as alternatives to radioactivity-based assays. However, fluorescence-based technologies are not devoid of pitfalls. One of the main problems is interference from autofluorescent compounds and the incidence of false-positives as exemplified here with a fluorescence polarization (FP)-based assay. Using the scintillation proximity assay as the in-house standard, we assessed several alternatives to radioactive methods, namely, the amplified luminescent proximity homeogeneous assay screen (ALPHAScreen™, Perkin-Elmer Life Sciences, Boston, MA), enzyme fragment complementation, FP, and nanofluidics-based fluorescence intensity. Data comparing the sensitivity, robustness, relative sensitivity to autofluorescent compounds, enzyme consumption, and relative costs of each assay for one common kinase are presented. Results obtained seem t...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    26
    References
    12
    Citations
    NaN
    KQI
    []