Application of liquid chromatography-turbo ion spray tandem mass spectrometry for quantitative analysis of a potent motilin receptor agonist, EM574, and its metabolites in human plasma.

1999 
Abstract A liquid chromatography–tandem mass spectrometry (LC–MS–MS) method for the simultaneous determination of a new potent motilin receptor agonist as erythromycin derivative, EM574 (erythromycin derivative), and its three metabolites, M-IV, M-V and M-VI, in human plasma was developed. The internal standards (I.S.s) used were deuterated EM574, M-IV and M-V. For the quantitation of M-VI, deuterated M-V was used. The analytes and I.S. were extracted from plasma samples with diethyl ether at neutral pH. A turbo ion spray interface was used as the ion source of LC–MS–MS, and the analysis was performed in the selected reaction monitoring mode. The lower quantitation limits for all the analytes were 0.05 ng/ml when 0.2 ml of plasma was used, and the standard curves were linear in the range 0.05 to 20 ng/ml. The method was precise; the intra- and inter-day precisions of the method were not more than 19.8% for all the analytes. The accuracy of the method was good with the deviations between added and calculated concentrations of each analyte being typically within ±11.2%.