Interlaboratory evaluation of milk fatty acid composition by using different GC operating conditions

2013 
Abstract The aim of this work was to test the effectiveness of an analytical protocol, including the calibration with a reference milk fat and the internal standard method, on milk fatty acid (FA) composition. The performances of both different types of stationary phases and column length (30–100 m), together with the elution order of some isomers, were verified. This latter was dependent not only on the polarity and length of the GC column, but also on the temperature gradient. The calibration of the instrument, using a natural reference material, affected the precision of the ISO 15885 standard in a positive way, lowering both the repeatability (from 0.5 to 0.15%, for the range of FA concentration varying from 1 to 5%; from 1 to 0.75% for the range of FA concentration varying from 5 to 10%) and the reproducibility (from 1 to 0.24% for the range of FA concentration varying from 1 to 5%; from 4 to 0.8% for the range of FA concentration varying from 5 to 10%). Moreover, the application of the response factors provided a correct and precise quantification of the short-chain FAs (4–8 carbon atoms). The protocol applied can be considered a useful example for further collaborative studies aiming to standardise the milk FA analysis with long polar GC columns.