Liquid Chromatography Quadrupole Time-Of-Flight Tandem Mass Spectrometry for Selective Determination of Usnic Acid and Application in Pharmacokinetic Study
2013
v/v) at a flow rate of 0.3 mL/min. A tandem mass spectrometric detection with an electrospray ionization (ESI) interface was conducted via collision induced dissociation (CID) under negative ionization mode. The MS/MS transitions monitored were m/z 343.0448 → m/z 313.2017 for usnic acid and m/z 153.1024 → m/z 136.2136 for protocatechuic acid (internal standard). The linear range was calculated to be 2.0-160.0 ng/mL with a detection limit of 3.0 pg/mL. The inter- and intra-day accuracy and precision were within ± 7.0%. Pharmacokinetic study showed that the apartment of usnic acid in vivo confirmed to be a two compartment open model. The method was fully valid and will probably be an alternative for pharmacokinetic study of usnic acid.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
18
References
4
Citations
NaN
KQI